Terminated

Jadak Barcode Scanner, Part No. 350587-32, 350587-41, 351377-32. Product Usage: An accessory for the Pyxis Anesthesia ES system and Pyxis Anesthesia system 4000 stations. The Jadak Barcode Scanner is used to scan medication labels during refill of the Pyxis Anesthesia ES system and Pyxis Anesthesia system 4000 stations and dispensing medication for patients.

FDA Recall Z-0814-2016 · Posted 2016-02-16

Recalling FirmCareFusion 303, Inc.
Recall StatusTerminated
Product CodeBRY
Date Initiated2016-01-15
Date Terminated2017-05-08
Product Quantity9,345 units

Reason for Recall

Customers reported issues when scanning medications with the scanners. When a user scans a medication using the affected barcode scanner, information on a different medication could be communicated to the Pyxis product.

Firm Action

BD sent an Urgent Medical Device Recall Notification letter dated January 15, 2015 to all affected customers The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete the enclosed Customer Response Card by selecting the desired deployment option and return the card to CareFusion. Customers with questions are instructed to contact the CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact CareFusion Customer Advocacy at (800) 836-7842. Customers with technical questions are instructed to contact CareFusion Technical Support at (800) 727-6102.

Root Cause

Device Design

Distribution Pattern

Worldwide:Distribution - US Nationwide and the countries of Saudi Arabia, Bahrain, Australia, United Arab Emirates, Qatar, Mexico, Guam and the Bahamas.