GE Healthcare-Americas Signa 3.0T EXCITE Whole Body Magnetic Resonance System

FDA Recall Z-0815-2007 · Posted 2007-05-18

Recall StatusTerminated
Product CodeLNH
Date Initiated2006-06-01
Date Terminated2008-02-24
Product Quantity30 units

Reason for Recall

To provide the 3.0T MRI systems users with the 8-channel brain coil the proper coil weighting software. The proper coil weighting factors for the eight-channel brain coil were not included in the software.

Firm Action

The corrective action is already completed and all customers have been notified via a personal visit to each site, to install a software patch beginning June, 2006.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, China, Finland, Germany, Israel, Italy, Korea, Spain, and Taiwan.