Transseptal Sheath Introducer Kits (under trade names: HeartSpan, Channel FX, Torflex, Braided Guiding Introducer Kit). The product is shipped 5 per carton. Intended to allow left heart catheterization procedure to occur through the right atrium.

FDA Recall Z-0815-2010 · Posted 2010-03-04

Recall StatusTerminated
Product CodeDYB
Date Initiated2010-02-01
Date Terminated2011-01-05
Product Quantity13,973 units

Reason for Recall

Radiopaque tip may fracture

Firm Action

Thomas Medical Products, A GE Healthcare Company, issued an "Urgent Medical Device Correction" letter dated February 1, 2010. Consignees were informed of the affected product then asked to cease distribution, return any product to firm and contact all sub-accounts. For further information, contact Thomas Medical Products Customer Service at 1-866-446-3003.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution -- United States (CA, MA and NJ), Canada, Germany, and the United Kingdom.