BD Vacutainer Push Button collection Set; Non-Pyrogenic, Sterile, Rx only; 23 G x 3/4 x 12" (0.6 x 19 mm x 305 mm) Becton Dickinson & Co., Franklin Lakes, NJ, USA Intended use: Blood collection
FDA Recall Z-0816-2011 · Posted 2010-12-27
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeFMI
Date Initiated2010-10-05
Date Terminated2012-06-06
Product Quantity4,223,200
Reason for Recall
There is the possibility for certain lots of the BD Vacutainer Push Button Blood Collection sets to have separation of the cannula from the hub.
Firm Action
BD Diagnostics sent "RECALL NOTIFICATION" separate letters October 1 and 5, 2010 and follow up letter dated October 8, 2010, as well as a letter dated December 1, 2010 to all distributors and customers commencing October 5, 2010 by federal express. The letters describe the product, the problem and the action to be taken by the customers. The customers were instructed to immediately examine their inventory for specific catalog and lot numbers, and return product from the lots following the instructions on the enclosed packing slip (BD will send replacement upon receipt of returned product). If you have any questions on returning and replacing product, please contact BD logistics at 1-866-380-0046.
If you have any further detail questions regarding this recall, please contact your BD Sales Consultant or BD Technical Services via email at www.BD.com/Vacutainer/Contact or via telephone at 1-800-631-0174.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution