Cloth electrodes with Silver-Carbon construction: P/N 202-9231, Cat No. 400-899, 2" x 2" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9241, Cat No. 400-895, 2" Round Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9381, Cat No. 400-889, 2" x 3.5" Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9501, Cat No. 202-9501, 1.25" Round Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8991, Cat Nos.: 400-877T and 400-899T, 2" x 2.25" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.
FDA Recall Z-0816-2013 · Posted 2013-02-14
Advertisement
Recalling FirmNAImco Inc dba Richmar Inc
Recall StatusTerminated
Product CodeGXY
Date Initiated2009-11-30
Date Terminated2013-09-03
Product Quantity25860 electrodes
Reason for Recall
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
Firm Action
Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
Distributed in the states of GA and TN.