Foam Electrodes with Carbon construction: P/N 201-1261, Cat Nos.: 201-126, 203-678, and 400-856, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1271, Cat Nos.: 201-127, 202-093, 203-483, 203-565, 203-673, and 400-873, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1281, Cat Nos.: 201-128, 202-277, 203-675, and 400-873, 2" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1291, Cat Nos.: 201-129, 203-677, and 400-882, 3" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1301, Cat Nos.: 201-130, 203-679, and 400-855, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1371, Cat Nos.: 201-137, 202-278, 203-484, 203-681, and 400-880, 2" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-5671, Cat No. 202-567, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.
FDA Recall Z-0817-2013 · Posted 2013-02-14
Advertisement
Recalling FirmNAImco Inc dba Richmar Inc
Recall StatusTerminated
Product CodeGXY
Date Initiated2009-11-30
Date Terminated2013-09-03
Product Quantity129400 electrodes
Reason for Recall
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
Firm Action
Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
Distributed in the states of GA and TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.