ARCHTECT System RS-232 Manual, product list number 06F71-04, for the ARCHITECT i2000 Processing Module and the ARCHITECT c8000 Processing Module. The product is manufactured and distributed by Abbott Laboratories Diagnostic Division, Irving, TX 75038
FDA Recall Z-0818-04 · Posted 2004-07-20
Advertisement
Recalling FirmAbbott Laboratories, Inc
Recall StatusTerminated
Product CodeJJE
Date Initiated2004-03-19
Date Terminated2005-11-03
Product Quantity101
Reason for Recall
When using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual.
Firm Action
Customers were notified by letter on March 23, 2004 to contact their LIS vendor and to reconfigure the ASTM Field 10.1.4 to allow a maximum character length of 20 characters.
Root Cause
Other
Distribution Pattern
Nationwide and to the following foreign countries: Chile, Canada, Germany, Taiwan, Australia, Signapore