Terminated

EB- Series Video Bronchoscopes, EB-1570K & EB-1970K. Intended to provide optical visualization of (via a video monitor), and therapeutic access to, the Pulmonary Tract.

FDA Recall Z-0818-2012 · Posted 2012-01-20

Recall StatusTerminated
Product CodeEOQ
Date Initiated2011-09-30
Date Terminated2012-11-27
Product Quantity18 units

Reason for Recall

The wrong silicone sealant was used and excess sealant was neglected to be removed during repair of the bronchoscope.

Firm Action

Pentax sent an "Urgent: Device Recall - Stop Use Immediately" letter/return response form to their customers on 9/30/2011. The letter identified the affected product and asked customers to immediately discontinue the use of their endoscope(s). Customers were to contact PENTAX at 1-800-736-8293 to schedule the return of the affected product(s). The reason for the recall was also explained, and the letter stated that there will be no charge for shipping, evaluation, or repair of the products. Customers can request loaner instruments, if necessary.

Root Cause

Employee error

Distribution Pattern

Nationwide Distribution.