VITROS Immunodiagnostic Products Troponin I Reagent Pack, REF/Catalog No. 194 9882. A Troponin Immunoassay Method, IVD. Ortho-Clinical Diagnostics, Amersham UK
FDA Recall Z-0820-2007 · Posted 2007-05-26
Advertisement
Recalling FirmOrtho-Clinical Diagnostics
Recall StatusTerminated
Product CodeMMI
Date Initiated2007-04-27
Date Terminated2012-09-07
Product QuantityLot 3151: 1,750 units, Lot 3170: 3,316 units
Reason for Recall
Quality control and patient results can be lower than expected when the true Troponin I value is less than or equal to 0.2 ng/mL which may result in the device failing to detect myocardial injury.
Firm Action
On 4/27/07, Ortho-Clinical Diagnostics in Rochester, NY, notified the end users of the recall of the product by phone, fax, and FedEx overnight letter. Clinical labs in possession of these lots have been contacted by OCD, and were instructed to (a) discontinue use of this product and (b) notify healthcare providers who ordered the test in recent weeks. Replacement product has been provided to clinical labs with the affected product lots. Clinical laboratories with questions may contact the company at 1-800-421-3311. On 4/27/07, OCD foreign affiliates were informed of this action by e-mail and instructed to inform their customers who received these reagent packs of this issue. Ortho-Clinical Diagnostics in Raritan, NJ, issued a nationwide press release on 5/04/07.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide-including USA, Canada, and Bermuda
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.