Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

EE- Series Video Naso Pharyngo Laryngo Scope, VNL-1570STK. Intended to provide optical visualization (via a video monitor) of the Upper Airway.

FDA Recall Z-0820-2012 · Posted 2012-01-20

Recall StatusTerminated
Product CodeEOX
Date Initiated2011-09-30
Date Terminated2012-11-27
Product Quantityone unit

Reason for Recall

The wrong silicone sealant was used and excess sealant was neglected to be removed during repair of the bronchoscope.

Firm Action

Pentax sent an "Urgent: Device Recall - Stop Use Immediately" letter/return response form to their customers on 9/30/2011. The letter identified the affected product and asked customers to immediately discontinue the use of their endoscope(s). Customers were to contact PENTAX at 1-800-736-8293 to schedule the return of the affected product(s). The reason for the recall was also explained, and the letter stated that there will be no charge for shipping, evaluation, or repair of the products. Customers can request loaner instruments, if necessary.

Root Cause

Employee error

Distribution Pattern

Nationwide Distribution.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑