Viterion TeleHealth Monitor Model Number V100, Product Codes 09377512 (US) and 04480218 (UK). Physiological patient monitor (without arrhythmia detection or alarms). Viterion V100, V100-BGM and V500 Telehealth Monitors transfer medical information.
FDA Recall Z-0820-2013 · Posted 2013-02-14
Advertisement
Recalling FirmViterion TeleHealthcare Llc
Recall StatusTerminated
Product CodeMWI
Date Initiated2012-11-15
Date Terminated2017-02-16
Product QuantityDomestic: 19,106 units; Foreign: 267 units
Reason for Recall
Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04 meters to healthcare professionals.
Firm Action
Bayer Healthcare sent "IMPORTANT RECALL INFORMATION" to all affected customers.The letter identified the product, problem, and actions to be taken by the customers. Contact Viterion Support at 1-800-866-0133 option 1 for questions relating to this issue.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution-USA (nationwide) and the United Kingdom.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.