Covidien Endo GIA Ultra Universal Standard Stapler Handle Ref: EGIAUSTND Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers when used with the Endo GIA curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.

FDA Recall Z-0820-2016 · Posted 2016-02-18

Recalling FirmMedtronic
Recall StatusTerminated
Product CodeGDW
Date Initiated2015-12-02
Date Terminated2019-07-18
Product Quantity1,102,166 units

Reason for Recall

Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use

Firm Action

Medtronic sent an Urgent Medical Device Recall letters dated December 2 2015 and an Updated Urgent Medical Device Recall letter dated January 13, 2016, to affected customers, via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473. For questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at (800) 882-5878.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US Nationwide and the countries of Canada, Albania, ALGERIA, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Greece, Guatemala, Honduras, Hungary, Iceland, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mauritius, Mexico, Monaco, Montenegro, Morocco, Namibia, Netherlands, Nicaragua, Nigeria, Norway, Panama, Paraguay, Peru, Poland, Portugal, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, UAE, Ukraine, United Kingdom, Uruguay, Venezuela, and Zimbabwe.