Terminated

Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3

FDA Recall Z-0821-2020 · Posted 2020-01-19

Recalling FirmIsopure Corp
Recall StatusTerminated
Product CodeKPO
Date Initiated2019-04-19
Date Terminated2020-05-29
Product Quantity15

Reason for Recall

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Firm Action

The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.

Root Cause

Nonconforming Material/Component

Distribution Pattern

The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.