BD Beaver Slit Blade 3.0mm Straight Part Number: 375521 The BD Beaver Slit Blade is used to perform the main incision through which instruments are inserted during cataract surgery.
FDA Recall Z-0822-2011 · Posted 2010-12-28
Advertisement
Recalling FirmBd Medical Systems
Recall StatusTerminated
Date Initiated2010-06-24
Date Terminated2012-06-29
Product Quantity524 pieces
Reason for Recall
Blades may contain defects at the blade tip or cutting edge.
Firm Action
BD Medical notified customers via letter dated 6/24/10 to discontinue use of product and remove from inventory. Customers were instructed to complete the customer response form and fax to BD Medical. BD Medical was to contact customers upon receipt of the customer response form to arrange for replacement.
Unused product was to be returned to:
BD Medical
Ophthalmic Systems
Attn: WT-12582
411 Waverly Oaks Road
Building 2, Suite 229
Watham, Massachusetts 02452-8402
For questions regarding this recall call 1-781-906-7952.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution