Terminated

Viterion TeleHealth Monitor Model Number V500; Product Codes 08832984 (US) and 02080468 (UK) - Physiological patient monitor (without arrhythmia detection or alarms). Viterion V100, V100-BGM and V500 Telehealth Monitors transfer medical information.

FDA Recall Z-0822-2013 · Posted 2013-02-14

Recall StatusTerminated
Product CodeMWI
Date Initiated2012-11-15
Date Terminated2017-02-16
Product QuantityDomestic - 296 units, Foreign - 73 units

Reason for Recall

Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04 meters to healthcare professionals.

Firm Action

Bayer Healthcare sent "IMPORTANT RECALL INFORMATION" to all affected customers.The letter identified the product, problem, and actions to be taken by the customers. Contact Viterion Support at 1-800-866-0133 option 1 for questions relating to this issue.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the United Kingdom.