Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)
FDA Recall Z-0823-2020 · Posted 2020-01-19
Advertisement
Recalling FirmIsopure Corp
Recall StatusTerminated
Product CodeN/A
Date Initiated2019-04-19
Date Terminated2020-05-29
Product Quantity8
Reason for Recall
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Firm Action
The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.
Root Cause
Nonconforming Material/Component
Distribution Pattern
The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.