Hobbs Medical 3 Prong Looped Retriever, a disposable device to be used by physician to remove foreign object or cauterize polyp specimens from the GI tract. REF: 5030 and 5030S
FDA Recall Z-0824-06 · Posted 2006-05-03
Advertisement
Recalling FirmHobbs Medical, Inc.
Recall StatusTerminated
Product CodeKOG
Date Initiated2006-03-15
Date Terminated2006-06-27
Product Quantity254 units
Reason for Recall
Diameter of device at the distal end would prevent it from passing through the endoscope's working channel.
Firm Action
Hobbs Medical notfied customers by Registered Letter on 3/15/06, requesting return of recalled product.
Root Cause
Process control
Distribution Pattern
AK, IA, KY, MI, NH, NC,UT, WI, Canada, and UK