Terminated

ACUSON SC2000 Power Input: 100-240VAC, 1600 Max 50/60Hz; Siemens Medical Solutions USA, Inc. 685 East Middlefield Road Mountain View, CA 94043 Ultrasound imaging system.

FDA Recall Z-0825-2014 · Posted 2014-01-24

Recall StatusTerminated
Product CodeIYN
Date Initiated2013-12-22
Date Terminated2015-03-05
Product Quantity271 units

Reason for Recall

The locking mechanism that is intended to hold the control panel in a fixed position can become loose and fail to lock the rotation of the control panel, making it unable to steer or control during transport.

Firm Action

Customer Safety Advisory Letters were sent to all affected users on December 22, 2013. Letters advised of the issues and the actions they should take until a Siemens Field Engineer will visit and upgrade the control panel.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Austria, Canada, Ecuador, Finland, France, Germany, India, Ireland, Israel, Italy, Japan, Mexico, P.R. China, Philippines, Portugal, Republic Korea, Russian Fed. Saudi Arabia, Serbia, South Africa, Spain, Suriname, Sweden, Taiwan, Turkey, U.A.E., and United Kingdom.