Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular surface of the femoral head in primary hip surgery and for the salvage of a failed previous hip surgery.

FDA Recall Z-0825-2015 · Posted 2014-12-17

Recall StatusTerminated
Product CodeLZO
Date Initiated2014-11-07
Date Terminated2015-03-24
Product Quantity19 units worldwide

Reason for Recall

Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to tissue tension and not the taper connection between the femoral head and stem. The investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the

Firm Action

On 11/7/2014, URGENT INFORMATION - RECALL NOTICE notifications were sent to the affected distributors, hospitals, and doctors with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospitals and doctors were notified via sales consultant visit. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For clinical questions from surgeons, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT). For questions about device recall information provided, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT).

Root Cause

Mixed-up of materials/components

Distribution Pattern

Distributed in the states of WV, AZ, and AL, and the countries of Germany, and Finland.