Terminated

Guidant CONTAK RENEWAL 4 (models H190, H195) CONTAK RENEWAL; 4 HE (model H197). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Cardiac resynchronization therapy defibrillators (CRT-D) provide ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with sudden cardiac death (SCD). Cardiac resynchronization therapy is for the treatment of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. Cardioversion/defibrillation therapies include a range of low-and high-energy shocks using either a biphasic or monophasic waveform.

FDA Recall Z-0826-06 · Posted 2006-05-02

Recalling FirmGuidant Corporation
Recall StatusTerminated
Product CodeNIK
Date Initiated2006-04-07
Date Terminated2007-04-01
Product Quantity20 devices

Reason for Recall

77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would likely render the device non-functional, resulting in no output and/or telemetry communications.

Firm Action

Guidant Sales Representatives are retrieving non-implanted devices April 10-12, 2006. If a device is implanted the physician is sent an Urgent Medical Device Safety Information & Corrective Action letter, April 11-14, 2006. This letter describes the issue, states that Guidant is retrieving all non-implanted devices and that Guidant is currently conducting tests on retrieved devices. Physicians will be advised once more information is available.

Root Cause

Other

Distribution Pattern

Nationwide. AZ, FL, GA, IL, IN, KY, MA, ME, MI, MN, NC, NE, NY, OH, PA, TN, TX, UT, VA, WI. OUS to include: Australia, Canada, Germany, Italy, Netherlands, Portugal, Slovenia, Spain.