Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left
FDA Recall Z-0826-2010 · Posted 2010-02-24
Advertisement
Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeJWH
Date Initiated2009-12-11
Date Terminated2010-09-24
Product Quantity17
Reason for Recall
Partial lot of product was not manufactured to specification. Central locking screw was missing threads. There is a potential for dissociation because the screw is not able to be used as a secondary fixation method.
Firm Action
Firm notified consignees by letter on 12/11/2009. The letter informs physicians that patients with implanted devices should counseled about the issue and should report any changes to their treating physician.
Root Cause
Process control
Distribution Pattern
TX, FL, GA, CA, AZ, Japan
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.