Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Stryker Aero-AL Impaction handle Catalog Number 48921007. Stryker Spine Manufactured by : Stryker Spine SAS France Used to facilitate insertion of the Aero-AL cage into the disc space.

FDA Recall Z-0828-2016 · Posted None

Recalling FirmStryker Spine
Recall StatusTerminated
Product CodeLXH
Date Initiated2015-12-18
Date Terminated2016-08-01
Product Quantity31 units

Reason for Recall

Reported jamming of the impaction handle when assembled to the Inserter guide at the threads.

Firm Action

Stryker Spine sent Urgent Product Removal notification letters to branches/agencies on 12/18/2015 via FedEx. The letters provided a description of the affected product, the issue, potential hazards and risk mitigations, and actions to be taken. Customers were asked to examine their inventory and hospital locations to identify the product and reconcile with the affected catalog and lot numbers provided. The Customer Response Form and Product Quantity Tracking Spreadsheet for Health Care Facilities should be completed and returned within five days of receipt of the notice to the e-mail or fax number provided. Questions can be directed to Nabil Riaz - Regulatory Compliance Specialist, Regulatory Compliance & Quality Systems at [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution

Official Source

View FDA Recall →