Stat Profile Critical Care Xpress CCX+ (R) Anaylzer Catalog NUmber: 37555 R= Refurbished

FDA Recall Z-0829-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeCHL
Date Initiated2004-03-30
Date Terminated2004-06-17
Product Quantity7units

Reason for Recall

Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems

Firm Action

Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.

Root Cause

Other

Distribution Pattern

CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany