2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

FDA Recall Z-0829-2018 · Posted 2018-02-20

Recall StatusTerminated
Product CodeKDI
Date Initiated2017-10-05
Date Terminated2021-07-28
Product Quantity15 machines

Reason for Recall

While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.

Firm Action

Fresenius Medical Care sent an Urgent Medical Device Correction letter dated April 24, 2017 to affected customers. the letter identified the affected product, problem and actions to be taken. For questions contact Technical Services at 1-800-227-2572.

Root Cause

Software Design Change

Distribution Pattern

US Nationwide Distribution