Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The Brilliance iCT and iCT SP are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
FDA Recall Z-0832-2014 · Posted 2014-01-27
Recalling FirmPhilips Medical Systems (Cleveland) Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2013-12-09
Date Terminated2014-05-29
Product Quantity7 units
Reason for Recall
Patient images exhibited ring artifacts.
Firm Action
The firm, Philips Healthcare, sent an "URGENT - Medical Device Correction" letter dated December 9, 2013 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following to avoid the occurrence of these specific ring artifacts:
1. Disable Automatic Scan Time for all Exam Cards
2. Leave Automatic Scan Time on during planning and prior to clicking Go
to initiate the scan, click the acquisition series of the scan and disable
Automatic Scan Time, then proceed with the scan.
3. Disable Edit Before Final Recon (Reconstruction) for all Exam Cards
If a ring artifact is present, clinical judgment should be used by the
customer/user to determine if the images acquired in the dataset contain enough information to make a diagnosis.
Philips will be releasing a Field Change Order (FCO) to correct the issue with software version release 4.1.1.19261. In addition, a Philips Field Service Engineer will contact you to schedule the software installation at your site.
If you need any further information or support concerning this issue, please
contact your local Philips representative or local Philips Healthcare office. For
North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Root Cause
Software design
Distribution Pattern
Worldwide Distribution: US (nationwide) including states of: FL, IN, NY, OR and PA; and internationally to: Australia and Israel.