Wills-Oglesby Percutaneous Gastrostomy Set Product Usage: Intended for the percutaneous placement of a catheter for gastroenteric feeding
FDA Recall Z-0832-2018 · Posted None
Recalling FirmCook Inc.
Recall StatusTerminated
Product CodeKNT
Date Initiated2017-10-19
Date Terminated2019-04-16
Product QuantityQuantities co-mingled into line 1
Reason for Recall
Cook has identified that the affected lots may contain wire guides that were loaded backwards into the wire guide holder.
Firm Action
On October 19, 2017 an URGENT MEDICAL DEVICE RECALL letter was mailed to customers requesting that they examine inventory for affected product, immediately quarantine and return product to recalling firm for credit. Questions or concerns can be directed to 1-800-457-4500 or 812-330-2235.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide Distribution