Terminated

GE eXplore Locus Micro CT cabinet x-ray system model RS9-80. CT Scanner for laboratory mice or other research.

FDA Recall Z-0833-2012 · Posted 2012-02-17

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeRCE
Date Initiated2011-01-31
Date Terminated2012-09-06
Product Quantity21

Reason for Recall

Product fails to comply with the federal performance standard for cabinet x-ray systems (Title 21 CFR 1020.40(c)(1)(i)). Specifically, the x-ray emission limit of 0.5 milliroentgen in one hour could be exceed. Highest emission measured was 3 mR/hr.

Firm Action

GE Healthcare will notify purchasers of the problem, the affected product, instructions for users, and specify the maximum emission rate measured. GE Healthcare will bring defective systems into compliance free of charge. This will be done via Field Modification Instructions IFMI) 11101 which is scheduled for deployment on March 1, 2011 with an estimated completion date of July 1, 2011.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

Worldwide Distribution