GE eXplore Locus Micro CT cabinet x-ray system model RS9-80. CT Scanner for laboratory mice or other research.
FDA Recall Z-0833-2012 · Posted 2012-02-17
Advertisement
Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeRCE
Date Initiated2011-01-31
Date Terminated2012-09-06
Product Quantity21
Reason for Recall
Product fails to comply with the federal performance standard for cabinet x-ray systems (Title 21 CFR 1020.40(c)(1)(i)). Specifically, the x-ray emission limit of 0.5 milliroentgen in one hour could be exceed. Highest emission measured was 3 mR/hr.
Firm Action
GE Healthcare will notify purchasers of the problem, the affected product, instructions for users, and specify the maximum emission rate measured. GE Healthcare will bring defective systems into compliance free of charge. This will be done via Field Modification Instructions IFMI) 11101 which is scheduled for deployment on March 1, 2011 with an estimated completion date of July 1, 2011.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Worldwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.