ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping Surface technology), 45MM WHITE VASCULAR/THIN, Product No. GST45W
FDA Recall Z-0834-2024 · Posted 2024-01-26
Advertisement
Recalling FirmEthicon Endo-Surgery Inc
Recall StatusOpen, Classified
Product CodeGDW
Date Initiated2023-12-11
Product Quantity22,616 (US), 4,476 (OUS)
Reason for Recall
There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.
Firm Action
On December 11, 2023, the firm notified affected consignees through URGENT VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) letters.
Customers were instructed to cease use and distribution of affected product lots.
Customers should examine their inventory immediately to determine if they have affected product. Affected product should be quarantined. Unused affected product should be returned.
If you require any assistance with returning product, please contact Sedgwick at 888-591-5087 and reference Event # 6925.
If you have additional questions regarding this voluntary medical device recall, please contact the Ethicon Resource Department at 1-877-ETHICON (1-877-384-4266). The Ethicon Resource Department is open Monday through Friday, 8:00 AM to 5:00 PM ET.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada and Brazil.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.