DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, REF T653NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

FDA Recall Z-0836-2014 · Posted 2014-01-27

Recall StatusTerminated
Product CodeILO
Date Initiated2013-09-17
Date Terminated2017-06-05
Product Quantity409 units

Reason for Recall

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Firm Action

DeRoyal sent a "URGENT VOLUNTARY RECALL DEROYAL HOT/COLD THERAPY UNIT" letter dated September 17, 2013. The letter identified the product, problem, and actions to be taken by the customers.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution.