Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.

FDA Recall Z-0836-2015 · Posted 2014-12-19

Recall StatusTerminated
Product CodeLHN
Date Initiated2014-11-07
Date Terminated2015-06-29
Product Quantity1

Reason for Recall

Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du

Firm Action

The firm, IBA, sent a " User Notice" dated November 7, 2014, to the sole affected site-UPHS Customer. The notice describes the product, problem and actions to be taken. The customer was instructed that the KV/kV source acquisition mode has been disabled and shall not be used for patient alignment until system is officially confirmed to be back within tolerances. IBA will change the kV/kV Single Source acquisition workflow configuration file to be linked to the correct geometrical calibration file and will update the kV/kV Single Source calibration file that includes the colinearity parameters, with the correct values. If you need assistance, contact the IBA Site Staff or the Operations manager at +32 10 475 997 or email: [email protected].

Root Cause

Software design (manufacturing process)

Distribution Pattern

US Distribution in PA only.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑