Terminated

ZIMMER UNI HIGH FLEX PRECOAT FEM prosthesis, knee, femorotibial, non-constrained, cemented, metal/polymer various sizes These devices are indicated for patients with: - Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis. - Previous tibial condyle or plateau fractures with loss of anatomy or function. - Varus or valgus deformities. - Revision of previous arthroplasty procedures. " These devices are indicated for cemented use only. " The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees."

FDA Recall Z-0836-2016 · Posted 2016-02-22

Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeHSX
Date Initiated2016-01-11
Date Terminated2017-05-23
Product Quantity32537

Reason for Recall

LDPE bag containing the implant adheres to the highly polished implant surface.

Firm Action

Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US (nationwide) and Internationally to ARGENTINA AUSTRALIA AUSTRIA BARBADOS BELGIUM BERMUDA BOLIVIA BRAZIL CARIBBEAN CHILE CHINA COLOMBIA COST RICA DENMARK DOMINICAN REPUBLIC DORAL FL ECUADOR EL SALVADOR ENGLAND FRANCE GERMANY GUATEMALA HONDURAS HONG KONG INDIA ISRAEL ITALY JAMAICA JAPAN KOREA MALAYSIA MEXICO NEW ZEALAND NICARAGUA PANAMA PERU SINGAPORE SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND VENEZUELA WEST INDIES Canary Islands Croatia Cyprus Czech Republic Canada