HORIZON Ez Arm (Part of the HORIZON TMS Therapy System.)

FDA Recall Z-0836-2019 · Posted None

Recall StatusOpen, Classified
Product CodeOBP
Date Initiated2018-10-25
Product Quantity27 (including 2 demos)

Reason for Recall

A stray strand of stainless-steel cable may protrude from the heat shrink sleeving on the Balance Cable of the Ez Arm.

Firm Action

On October 26, 2018, the firm sent URGENT FIELD SAFETY NOTICE, Medical Device Recall letters to their consignees to check their devices for this problem.

Root Cause

Nonconforming Material/Component

Distribution Pattern

CA, MA, MN, PA, TX, VA, WI, and WA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.