Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.
FDA Recall Z-0836-2025 · Posted 2025-01-10
Advertisement
Recalling FirmBiocartis Nv
Recall StatusOpen, Classified
Product CodeOOI
Date Initiated2024-12-10
Product Quantity8 units
Reason for Recall
False-positive MSI-H results generated by the Idylla MSI Test.
Firm Action
Biocartis issued an URGENT - Medical Device Correction notice to its consignees on 12/10/2024 via email. The notice provided the description of the problem, its impact, and requested the following actions be taken:
1. Until your Instrument is serviced, please move forward with the option that applies to your situation:
o In case you have another Idylla Instrument available that is not affected, use that
Instrument to run the Idylla MSI Test and discontinue running the Idylla MSI Test on the
listed Instrument(s).
o No new Idylla MSI testing should be performed using the affected Idylla Instruments.
o For results already obtained using an affected Idylla Instrument, the following apply:
o MSS results are VALID and may be reported.
o MSI-H (Microsatellite Instability High) results with six or seven positive identified
biomarkers (as depicted in the result report) are VALID and may be reported.
o MSI-H results other than those with six or seven positive identified biomarkers are
INVALID and should NOT be reported or used to manage patient health. Patient samples with these results should be re-tested on an Idylla Instrument that is NOT affected.
2. If you are not the Idylla user, please forward this notice to the Idylla end user or Idylla
supervisor and provide Biocartis with the correct contact details for our record keeping.
3. Completion of the Acknowledgement of Receipt (Appendix 2) is required for compliance. Please complete and sign the attached Acknowledgement of Receipt form by December 17, 2024, and send it to [email protected].
Root Cause
Environmental control
Distribution Pattern
US: AL, NJ, TX, OH, CA
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.