Atlantis PV Peripheral Imaging Catheter, model number H749364560, catalog number 36456, manufactured by Boston Scientific Corporation, Fremont, CA. Medical device, intended for ultrasound examination of peripheral pathology. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.

FDA Recall Z-0837-2009 · Posted 2009-01-28

Recall StatusTerminated
Product CodeITX
Date Initiated2008-10-09
Date Terminated2009-07-27
Product Quantity1033 units

Reason for Recall

Product is mislabeled as having a shelf life of 3 years, where the actual shelf life is 1 year.

Firm Action

Recall initiated on October 6, 2008, and notification letters sent on October 8, 2008. Letters were sent via FedEx. Responses are being tracked along with delivery verification. For additional information contact Boston Scientific Corporation at 1-508-652-5594.

Root Cause

Labeling mix-ups

Distribution Pattern

Worldwide Distribution --- USA including Puerto Rico and country of Panama