Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets
FDA Recall Z-0837-2011 · Posted 2011-01-10
Advertisement
Recalling FirmB Braun Medical, Inc
Recall StatusTerminated
Product CodeFRN
Date Initiated2010-08-24
Date Terminated2014-05-29
Product Quantity2,107 Pumps (582 Pumps to Winchester)
Reason for Recall
Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.
Firm Action
B. Braun Medical Inc. notified consignee by letter on August 24, 2010.
Root Cause
Component design/selection
Distribution Pattern
Virginia and New York