Regard Custom Procedure Kits containing BD Luer Lok Access Devices
FDA Recall Z-0837-2022 · Posted None
Advertisement
Recalling FirmROi CPS LLC
Recall StatusTerminated
Product CodeLRS
Date Initiated2022-02-24
Date Terminated2022-04-25
Product Quantity1750 kits
Reason for Recall
Recalled devices were packaged in convenience kits.
Firm Action
A customer letter dated February 24, 2022 was issued to the impacted consignee. The letter instructs the consignee to perform a microscopy to confirm an elevated RBC count in a urine sample as the impacted component may shed contaminant particles that are identified as RBCs when used in conjunction with the Sysmex UF-1000i or UX-2000 analyzers.
A reply form should be completed and returned to ROi confirming receipt and understanding of the notification.
End users should retain a copy of the BD notice and post near the analyzer as a visual reminder of the potential issue with the BD product.
Root Cause
Material/Component Contamination
Distribution Pattern
US Nationwide distribution in the state of Missouri.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.