Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Terragene Bionova PCD (PCD222-C)

FDA Recall Z-0837-2024 · Posted 2024-01-26

Recalling FirmTERRAGENE S.A.
Recall StatusOpen, Classified
Product CodeFRC
Date Initiated2023-12-07
Product Quantity80 stripes

Reason for Recall

Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products.

Firm Action

On 12/07/2023, the firm notified via email its initial importers/distributor a "Voluntary Recall Notification " that contained a "COMMUNICATION TO CUSTOMER VOLUNTARY RECALL OF PRODUCT/URGENT: MEDICAL DEVICE RECALL" Letter informing them of the discrepancies identified in the IFUs of some products that mentions extended cycles that may mislead users to think that these cycles are part of the FDA-cleared indications for use of the product. Initial Importers/Distributors are instructed to send the following information to Terragene: -Quantity in stock -Quantity distributed -Consignees On or about 12/22/2023, the firm initiated distribution, via email, an "URGENT: MEDICAL DEVICE RECALL" Letter to customers (identified by their Initial Imports/Distributors) informing them of the Recall issue (discrepancy identified in the IFU) provide an updated IFU (version 4) correcting the issue. Customers are instructed to print the updated IFU and attach it to their remaining inventory, and to discard the older IFU versions (version 1 thru 3). For questions or concerns, contact [email protected] or www.terragene.com , or +1-844-837-7243 Monday thru Friday 6:00 AM to 3:00 PM, Eastern Time.

Root Cause

Labeling False and Misleading

Distribution Pattern

U.S. Nationwide distribution in the states of GA, IN, MI, MT, NY, PA, TX and VA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑