DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Foot/Ankle Blanket NS, REF T656NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
FDA Recall Z-0839-2014 · Posted 2014-01-27
Advertisement
Recalling FirmDeRoyal Industries Inc
Recall StatusTerminated
Product CodeILO
Date Initiated2013-09-17
Date Terminated2017-06-05
Product Quantity25 units
Reason for Recall
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Firm Action
DeRoyal sent a "URGENT VOLUNTARY RECALL DEROYAL HOT/COLD THERAPY UNIT" letter dated September 17, 2013. The letter identified the product, problem, and actions to be taken by the customers.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution.