Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904
FDA Recall Z-0839-2019 · Posted None
Advertisement
Recalling FirmRandox Laboratories Ltd.
Recall StatusTerminated
Product CodeCJY
Date Initiated2018-12-04
Date Terminated2020-06-18
Product Quantity225 kits
Reason for Recall
Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.
Firm Action
Randox the US distributor contacted customers by email on 12/4/18 to advise of problem and provided an update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments.The instrument testing order should be reviewed in line with the updated technical bulletin. A revised letter issued 12 Feb 2019 to advise
"reagent would be observed as inconsistencies in Quality control
recovery which may lead to a delay in running patient samples or erroneous elevated / depressed test results".
Root Cause
Device Design
Distribution Pattern
WV, PR