Smith & Nephew, 420XL Xenon Light Source, 180 Watt; for endoscopic use. Catalog/Part #72200510; manufactured for Smith & Nephew Endoscopy, Andover, MA 01810.

FDA Recall Z-0840-2007 · Posted 2007-06-07

Recall StatusTerminated
Product CodeFFS
Date Initiated2007-04-09
Date Terminated2008-09-02
Product Quantity67 units.

Reason for Recall

Several units were manufactured using a power supply that contains a 1 amp fuse where a 5 amp fuse is required. When the unit's power supply draws more that 1A, the fuse will fail causing the light to cease operating. The problem can occur anytime: from the initial turning on (powering up), to anytime during use.

Firm Action

Notification sent 04/09/07 via FedEx to distribution centers, sales reps and user facilities. Users asked to send light source to Smith & Nephew GMBH Service Center in Germany for repair (contact Ute Dieling at +49 (0) 7462-208-313).

Root Cause

Nonconforming Material/Component

Distribution Pattern

Product sent to distributors in Argentina, Austria, Chile, Denmark, Dubai, France, Greece, Hong Kong, India, Italy, Japan, Korea, Malaysia, Norway, South Africa, Spain, Sweden and United Kingdom. No US distribution.