Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Completed

Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC71V; 2) WC71V-DEMO; 3) WC71V-R; used in conjunction with Warming Blankets and Mattresses, veterinary use.

FDA Recall Z-0842-2025 · Posted 2025-01-10

Recall StatusCompleted
Product CodeDWJ
Date Initiated2022-03-14
Product Quantity2271 units

Reason for Recall

HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".

Firm Action

Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.

Root Cause

Device Design

Distribution Pattern

Worldwide

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑