Simplastin HTF Reagents, 20ml HO, ( and 6 ml HO) bioMerieux, Inc., Durham, NC 27704
FDA Recall Z-0843-05 · Posted 2005-05-18
Advertisement
Recalling FirmBIOMERIEUX, INC.
Recall StatusTerminated
Product CodeGJS
Date Initiated2005-03-07
Date Terminated2006-10-16
Product Quantity19748 kits
Reason for Recall
The ISI labeling for Simplastin HTF on the Coag-a-Mate MAX and Thrombolyzer instrument platforms is not correct. The Simplastin HTF ISI assignment was mis-assigned and the published values are higher than expected for the listed instrument platforms.
Firm Action
Consignees were notified by letter on/about March 7, 2005.
Root Cause
Other
Distribution Pattern
Nationwide, Canada, Colombia, India, Philippines, Spain, Thailand