Centered Glenosphere w/Screw, CoCr/TA6V/TiN 10-degrees Tilt 32 mm, REF: 115-3210
FDA Recall Z-0843-2024 · Posted 2024-01-29
Advertisement
Recalling FirmFX SHOULDER
Recall StatusOpen, Classified
Product CodePHX
Date Initiated2023-08-08
Product Quantity4
Reason for Recall
Centered glenosphere box may contain an eccentric glenosphere.
Firm Action
On 8/8/23, the firm telephoned the customer and informed them of the recall. The customer was asked to quarantine the device, which was subsequently collected by the firm.
Root Cause
Labeling mix-ups
Distribution Pattern
US Nationwide distribution in the state of NJ.