Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC77V; 2) WC77V-DEMO; 3) WC77V-R; used in conjunction with Warming Blankets and Mattresses, veterinary use.
FDA Recall Z-0843-2025 · Posted 2025-01-10
Advertisement
Recalling FirmAugustine Temperature Management, LLC
Recall StatusCompleted
Product CodeDWJ
Date Initiated2022-03-14
Product Quantity1113 units
Reason for Recall
HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
Firm Action
Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.
Root Cause
Device Design
Distribution Pattern
Worldwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.