Terminated

AEROSET Analyzer

FDA Recall Z-0844-03 · Posted 2003-05-20

Recall StatusTerminated
Product CodeJJE
Date Initiated2002-08-15
Date Terminated2004-04-07
Product Quantity578

Reason for Recall

Design of high concentration waste line causes formation of a precipitate in the waste line causing overflow of waste.

Firm Action

The firm sent Technical Service Bulletins dated August 15, 2002 to consignees. Field Representatives are correcting the device in the field. A customer letter was released on August 20, 2002 to provide customers actions to take to prevent the occurrance.

Root Cause

Other

Distribution Pattern

Nationwide and to the following foreign countries: Mexico, Venezuela, Columbia, Argentina, Brazil, Puerto Rico, Dominican Republic, Canada, Germany, Hong Kong, New Zealand, and China