INFX-8000V Fluoroscopic X-Ray Systems; INFINIX CF-I/BP PLUS CARDIAC SYSTEM

FDA Recall Z-0844-2018 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2017-12-29
Date Terminated2023-11-29
Product Quantity29 units

Reason for Recall

During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.

Firm Action

A recall notification letter was sent to affected consignees on 12/29/2017. The following information was included in the notification letter: Should this problem occur prior to the above corrective action and in the middle of a clinical case, you will be able continue the case tentatively. The x-ray emissions will be still available and the inhibited ceiling support device can be overridden by pressing the override switch on the left side of the operation console. If any abnormalities are found during use, please stop using your system and contact your Toshiba service representative. Please share this information with all users and reviewing radiologist as well as clinical engineering and the biomedical group at your facility.

Root Cause

Process control

Distribution Pattern

United States