Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/30 cm, sterile, Cook Urological, Inc., Spencer, IN; Ref EUSCFS-060030-10-13; GPN REF G18121. Ureteral stent.

FDA Recall Z-0845-2009 · Posted 2009-01-23

Recall StatusTerminated
Product CodeFAD
Date Initiated2008-07-31
Date Terminated2009-11-27
Product Quantity1739 of all products.

Reason for Recall

The bond may break where the two materials meet, resulting in separation of the stent.

Firm Action

Consignees were notified via "Urgent: Recall Notice" letter dated 7/31/08. The letter described the failure, instructed that the products be isolated and returned to the firm. For additional information, contact Cook Urological at 1-812-339-2235, ext. 7241.

Root Cause

Process control

Distribution Pattern

Nationwide, Australia, Canada, Colombia, Hong Kong, India, Ireland, Israel, Japan, Malaysia, Mexico, South Korea, Taiwan, Venezuela and Viet Nam