Terminated

Computed Tomography X-Ray Systems (Brilliance 64, Model Number: 728231; Brilliance Big Bore, Model Number: 728243; Brilliance iCT, Model Number: 728306; Brilliance iCT SP, Model Number: 728311; and Ingenuity CT, Model Number: 728326); Mfr. by Philips Medical Systems; Cleveland, OH. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

FDA Recall Z-0845-2012 · Posted 2012-01-24

Recall StatusTerminated
Product CodeJAK
Date Initiated2011-10-07
Date Terminated2013-06-18
Product Quantity548 units

Reason for Recall

The Common Image Reconstruction System (CIRS) on the affected Brilliance Systems encounters an error and requires reboot of the system. The unit failed to initialize axial scan due to timeout that occurred waiting for "READY" state

Firm Action

On 10/10/2011 the firm sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected products, the problem, the hazard involved, the action to be taken by the customer, and actions planned by Philips. Customers were to reboot their systems if the two possible scenarios occurred. Philips Healthcare plans on installing the update through a Field Change Order free of charge. For further information or support concerning the issue, contact a local Philips representative or local Philips Healthcare office. In North America and Canada, contact Customer Care Solutions Center at 1-800-722-9377, option 5.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- USA, including the states of: AL, AZ, CA, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA & WV and the countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BANGLADESH, BELGIUM, Brazil, CHINA, COLUMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KOREA, LUXEMBOURG, MALAYSIA, MALI, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PHILIPPINES, POLAND, QATAR, RUSSIA, SINGAPORE, SLOVENIA, SOUTH KOREA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN & VIETNAM.