InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908

FDA Recall Z-0845-2014 · Posted 2014-01-27

Recalling FirmPhilips Healthcare
Recall StatusTerminated
Product CodeNCX
Date Initiated2014-01-13
Date Terminated2014-02-13
Product Quantity50

Reason for Recall

Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.

Firm Action

Philips Healthcare Recall Notice for this proactive performance correction was provided verbally to hospital biomedical engineers. Philips product support engineering scheduled and conducted the update on customer devices to the most current MCB configuration.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US (Nationwide) and Internationally to: Austria, France, Germany The Netherlands, Spain, Sweden, and Switzerland.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.