Terminated

Altaire, Magnetic Resonance Imaging Device.

FDA Recall Z-0846-05 · Posted 2005-05-17

Recall StatusTerminated
Product CodeLNH
Date Initiated2004-11-18
Date Terminated2008-01-04
Product Quantity189 devices

Reason for Recall

The absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor.

Firm Action

The firm implemented site inspections of customers beginning on 11/19/04 which have been completed. Customers that had the suspect component had it replaced.

Root Cause

Other

Distribution Pattern

Direct accounts located in AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.